| (618 intermediate revisions by 14 users not shown) |
| Line 1: |
Line 1: |
| __TOC__
| | #REDIRECT [[Main Page]] |
| | |
| | |
| ==OpenVMS Commands==
| |
| | |
| ===HowTo: Edit Scanco Submit Command Symbol===
| |
| | |
| In directory <code>DISK2:[MICROCT]MICROCT_SETUP_USER.COM</code> add the following line:
| |
| <pre>
| |
| $ SCANCO_SUBMIT_COMMAND :== SUBMIT/NOPRINT/QUE=SYS$SLOW /NOTIFY/LOG=SYS$SCRATCH:
| |
| </pre>
| |
| Then end session and log in again.
| |
| | |
| ===HowTo: Start a 3D evaulation from the terminal===
| |
| | |
| Warning: Do not change uct_evaulation_v4.com and ipl_fixed_batch_v4.com in .eval, but only work on your own file (i.e., eval_steve).
| |
| First, an explanation of how the normal evaluation runs by hitting the 'Start 3d' button.
| |
| $ scanco_submit_command ue:uct_evaluation_v4.com/para=(xxx,yyy,zzz,…)
| |
| There are 8 parameters:
| |
| *P1: number of print outs of our results sheet
| |
| *P2: the ISQ filename
| |
| *P3-5: start x (pos x), y, z
| |
| *P6-8: dim x, y, z
| |
| The parameters p3-p8 come from the white bounding box. The problem is that when you start from the terminal you do not have these. This problem can be overcome by using the isq_to_aim_gobjbox.exe and then give the relevant parameters. You can
| |
| $ run um: isq_to_aim_gobjbox.exe
| |
| to see what it's asking for.
| |
| By using gobjbox you do not need p3-p8. Therefore you can start the evaluation by giving:
| |
| $ scanco_submit_command ue:uct_evaluation_special_josh.com –/para=(1,dk0:[microct.data.0000275.0000245]c0000332.ISQ)
| |
| Note that you can make uct_evaluation_special_josh.com such that there is no need for P1 (change p2 to p1 in the script, etc).
| |
| [Alternatively to using the gobjbox you could use the normal isq_to_aim, but hard code the dim and pos.]
| |
| Then call whatever IPL files you want to call. Look into fixed_batch_v4 to see how we turn on the database always on the first line of every IPL part, and how we calculate parameters.
| |